fp: to execute a scan
fp -bcv BCVi BCVf -nstep n
(n is the number of steps of the scan)
fp -bcv BCVi BCVf -step step
(step is the step width of the scan)
fp -wave Wi Wf -nstep n
fp -wave Wi Wf -step step
fp -wave Wc -1 -nstep n
(to make a n-step scan around the value Wc, with a step size set by setgrsampling)
fp -lamb Wc
(to make an
image at the Wc wavelength)
setcalib: to adjust the wavelength calibration
setcalib lambda BCV
setparal: to adjust the parallelism of the FP
setparal BCVX BCVY
setgrsampling: to adjust the default step size of a
scan
setgrsampling stepsize
grifin: to set the approriate focus filter values when the FP is
in the beam (cannot be typed in Xterm with a bash shell)
grifin
grifout: to set the appropriate focus filter values when the
FP is out of the beam (cannot be typed in Xterm with a bash shell)
grifout
grifsetup: to set the desired setup of the Gumball lamps
grifsetup cont (to
light on all the continuum lamps)
grifsetup argon (to light
on all the argon lamps)
grifsetup neon (to
light on all the neon lamps, if any)
grifsetup krypton (to light on all krypton
lamps, if any)
grifsetup clean (to light
off all lamps)
grifsetup object (to be execute
before making a scan on an object)
/h/grif/.,grifh.par: in this file are stored the values selected
by the aforementioned commands (setcalib, setparal, setgrsampling, grifin/out,
grifsetup)
griffp: to send BCV to the CS 100 or reset the iotechs
griffp -z BCVZ (set
BCVZ on the Z channel)
griffp –xy BCVX BCVY (set
BCVX on the X channel and BCVY on the Y channel)
griffp -i (reset
the iotechs)
griffp -r (reset
the iotechs)